Features
- 6 1/2 Digit measurement resolution
- Dual measurement for Voltage and Current
- AC&DC Current measurement range:30mA/300mA/3A/30A/300A(Max)
AC Voltage measurement range:200mV/2V/20V/200V/600V(Max)
DC Voltage measurement range:200mV/2V/20V/200V/1000V(Max)
- Auto zero circuit
- Standard Interfaces: USB, GPIB
Description:
PCS-1000 is a High-Precision D.C. and A.C. Current Shunt Meter which carries built-in current shunts and high-accuracy current measurement circuits. PCS-1000 not only measures current in five current levels ─300A, 30A, 3A, 300mA and 30mA, but also measures voltage at 6 1/2-digit resolution simultaneously. This unique feature puts PCS-1000 in an unprecedented position among all the shunt meters available in the market today. After a long time operation, PCS-1000 is able to maintain high measurement accuracy in spite of the temperature variation. This is due to its core design based on the high stability and low temperature coefficient shunts. PCS-1000, a verification-lab level current and voltage measurement equipment, is suitable for the R&D, quality verification and testing jobs of all power products such as power supply, communications system power, server power and electronic load. The maximum voltage/current measurement ranges of PCS-1000, including DC 1000V / AC 600V / AC 300A / DC 300A, meet all current and voltage measurement challenges of every power industry.
Specification:
DC Voltage Characteristics | ||||
Range | 23℃±5℃(1 Year) | Temperature Cofficient / ℃ | ||
200.0000mV | 0.0050+0.0035 | 0.0005+0.0005 | ||
2.000000V | 0.0050+0.0010 | 0.0005+0.0001 | ||
20.00000V | 0.0050+0.0010 | 0.0005+0.0001 | ||
200.0000V | 0.0050+0.0010 | 0.0005+0.0001 | ||
1000.000V | 0.0050+0.0020 | 0.0005+0.0001 | ||
Accuracy | ±(% of reading +% of Range); Voltage input Resistance:10MΩ for all DC Voltage ranges | |||
DC Current Characteristics | ||||
Range | Burden Voltage | 23℃±5℃(1 Year) | Temperature Cofficient / ℃ | |
30.00000mA | <0.4V | 0.01+0.005 | 0.001+0.002 | |
300.0000mA | <0.5V | 0.01+0.005 | 0.001+0.002 | |
3.000000A | <0.8V | 0.01+0.005 | 0.001+0.002 | |
30.00000A | <0.8V | 0.01+0.005 | 0.001+0.002 | |
300.0000A | <0.8V | 0.02+0.005 | 0.001+0.002 | |
Accuracy | ±(% of reading +% of Range) | |||
True RMS AC Voltage Characteristics | ||||
Range | Frequency | 23℃±5℃(1 Year) | Temperature Cofficient / ℃ | |
200.0000mV | 45Hz-2KHz 2KHz-10kHz 10kHz-20kHz |
0.5+0.05 1.0+0.05 2.0+0.10 |
0.005+0.005 | |
2.000000V | 0.005+0.005 | |||
20.00000V | 0.005+0.005 | |||
200.0000V | 0.005+0.005 | |||
600.000V | 0.005+0.005 | |||
Accuracy | ±(% of reading +% of Range) ℃ | |||
True RMS AC Current(AC+DC) Characteristics | ||||
30.00000mA | 45Hz-2KHz 2KHz-10kHz |
0.5+0.05 1.0+0.05 |
0.03+0.006 | |
300.0000mA | 0.03+0.006 | |||
3.000000A | 0.03+0.006 | |||
30.00000A | 45Hz-400Hz | 0.5+0.05 | 0.03+0.006 | |
300.0000A | 0.03+0.006 | |||
Accuracy | ±(% of reading +% of Range) | |||
Current Monitor Accuracy | ||||
Range | Shunt Value | DC Accuracy | AC Accuracy ≦ 400Hz | Max Input DC/ AC RMS |
10Ω | 0.01% | 0.1% | 30mA | |
1Ω | 0.01% | 0.1% | 300mA | |
0.1Ω | 0.01% | 0.1% | 3A | |
0.01Ω | 0.01% | 0.1% | 30A | |
0.001Ω | 0.02% | 0.1% | 300A | |
General | ||||
Power Supply | 100V/120V/220V/240V±10% | |||
Power Line Frequency | 50/60Hz | |||
Operating Environment | Full Accuracy for 0℃ to 50℃, Full accuracy to 80% R.H. at 40℃ | |||
Storage Environment | -40℃ to 70℃ | |||
Power Consumprion | Max 35VA | |||
Dimensions | 210 mm (W) x 80mm(H) x 390 mm(D) | |||
Weight | Approx 6kg |
Accessories :
Standard Accessories:
User Manual x 1
AC Power Cord x 1
GTL-105A x 1 Alligator Clip Test Lead (3A Max)
GTL-207 x 1 Banana Plug Test Lead
GTL-240 x 1 USB Cable
PCS-001 x 1 Basic Accessory Kit
Optional Accessories:
GRA-419-J Rack Mount Adapter (JIS)
GRA-419-E Rack Mount Adapter (EIA)
Ordering Information:
PCS-1000 Precision Current Shunt Meter
Downlod :
Marcom Tool
Description
DATASHEET 2785 K PCS-1000 DATASHEET EN VA 2014/5/13
Manual
Description
User 1987 K PCS-1000 User Manual EN VA 2014/5/13